| Measuring Parameter | Turbidity | 
| Highlights | Light weight, IP67 waterproof meter | 
| Principle | Nephelometric non-ratio / ISO 7027 and DIN 27027 compliant | 
| Range | 0 to 2000 NTU | 
| Automatic Range Selection | 0.01 to 19.99 NTU;
 20.0 to 99.9 NTU; 100 to 2000 NTU  | 
| Resolution | 0.01 NTU (0 to 19.99 NTU);
 0.1 NTU (20.0 to 99.9 NTU); 1 NTU (100 to 2000 NTU)  | 
| Accuracy | < 0.1 NTU for 0.02 NTU standard;
 ±2 % of reading ±1 digit for 0 to 500 NTU; ±3 % of reading ±1 digit for 501 to 2000 NTU  | 
| Repeatability | < or = ±1 % of reading | 
| Calibration | Up to 4 points | 
| Calibration Standards | 0.02 NTU, 20.0 NTU, 100 NTU, 800 NTU | 
| Response Time | < 6 secs for full step change | 
| Sample Volume | 10 mL | 
| Light Source | Infrared-emitting diode (850 nm wavelength) | 
| Operating Temperature Range | 0 to 50 ºC | 
| Sample Temperature Range | 0 to 50 ºC | 
| Enclosure Type & Rating | ABS plastic & IP67 rated | 
| Power | 4 x 1.5 V ‘AAA’ alkaline batteries (> 1200 readings) | 
| Certifications/Compliance | FCC part 15 class A, TUV 3 in 1, CE, RoHS | 
| Dimensions (LxWxH) | Meter: 15.5 x 6.8 x 4.6 cm
 Boxed: 35.5 x 16.5 x 10.5 cm  | 
| Weight | Meter: 200 g
 Boxed: 1,160 g  | 
								