| Wavelength Range | 190 to 1,100nm |
| Spectral Bandwidth | 1nm (190 to 1,100nm) |
| Wavelength Display | 0.1 nm Increments |
| Wavelength Setting | 0.1Â Â nm Increments
(1 nm increments when setting scanning range) |
| Wavelength Accuracy | ±0.1 nm at D2 peak 656.1 nm,
± 0.3 nm for entire range |
| Wavelength Repeatability | – ±0.1 nm |
| Wavelength Slew Rate | About 14,500 nm/min |
| Wavelength scanning speed | 3,000 to 2 nm/min
29,000 nm/min when survey scanning |
| lamp interchange wavelength | Automatic interchange linked to wavelength. The interchange wavelength can be set freely in the range of 295 to 364 nm (0.1 nm increments). |
| Stray light | Less than 0.02% at 220 nm (Nal)
Less than 0.02% at 340 nm (NaNO2) Less than 0.5% at 198 nm (KCl) |
| Photometric system | Double beam optics |
| Photometric range | Absorbance: -4 to 4 Abs
Transmittance: 0% to 400% |
| Photometric accuracy | ±0.002 Abs at 0.5 Abs
± 0.004 Abs at 1.0 Abs ±0.006 Abs at 2.0 Abs (measured using NIST930D/NIST1930 or equivalent) |
| Photometric repeatability | Less than ±0.0002 Abs at 0.5 Abs
Less than ±0.0002 Abs at 1 Abs Less than ± 0.001 Abs at 2 Abs |
| Baseline stability | Less than 0.0003 Abs/Hr (700 nm, one hour after light source turned ON) |
| Baseline flatness | Less than ±0.0006 Abs (1,100 to 190 nm, one hour after light source turned ON) |
| Noise level | Less than 0.00005 Abs (700 nm) |
| Light source | 20-W halogen lamp and deuterium lamp Built-in light source auto position adjustment |
| Monochromator | LO-RAY-LIGH grade blazed holographic grating in Czerny-Turner mounting |
| Detector | Silicon photodiode |
| Sample compartment | Internal dimensions: W110 x D250 x H115 mm Distance between light beams: 100mm |
| Power requirements | AC100,120,220,230,240 V,
50/60 Hz, 140 VA |
| Environmental requirements | Temp: 15°C to 35°C
Humidity: 30% to 80% (without condensation; 70% max. at 30°C or higher) |
| Dimensions | W450 x D501 x H244 mm |
| Weight | 16.6 kg |
| Output device | USB memory (optional)
Data files saved in text format or UVPC format UVPC – format files can be read directly by UVProbe and LabSolutions UV-Vis |
| PC compatibility | UVProbe software (standard)
LabSolutions UV-Vis software (optional) External control possible via USB |
| Display | 24 – bit color touch screen |
| Supported languages | Japanese, English, Chinese, Spanish (Mexico), Portuguese (Brazil) |



























































Reviews